Stereotactic Body Radiation Therapy (SBRT) in prostate cancer in the presence of hip prosthesis – is it a contraindication? A narrative review

Abstract

Hip replacement is a common orthopedic surgery in the aging population. With the rising incidence of prostate cancer, metallic hip prosthetics can cause considerable beam hardening and streak artifacts, leading to difficulty in identifying the target volumes and planning process for radiation treatment. The growing use of Stereotactic Body Radiation Therapy (SBRT) to treat prostate cancer is now well established. However, the use of this treatment modality in the presence of a hip prosthesis is poorly understood. There is enough literature on planning for external beam radiation treatment without any difficulties in the presence of hip prosthesis with conventional or Hypofractionated treatment. However, there is a shortage of literature on the impact of the prosthesis in SBRT planning, and there is a need for further understanding and measures to mitigate the obstacles in planning for SBRT in the presence of hip prosthesis. We present our review of the intricacies that need to be understood while considering SBRT in the presence of hip prostheses in prostate cancer treatment.

Peer Review reports

Introduction

Stereotactic Body Radiation Therapy (SBRT) has evolved significantly over the last two decades in the management of prostate cancer [1]. However, there are scenarios where this method of treatment delivery is contraindicated [2]. The presence of hip implants has been considered a hindrance in treatment planning and sometimes can be viewed as a relative contraindication for consideration of SBRT treatment. Considering the age demographics of prostate cancer, hip replacements are prevalent in the elderly patient population [3], leading to challenges in delivering the adequate radiation treatment for prostate cancer. There has also been a note in the increase in the number of hip replacements being done over the years [4] from 210 per 100 000 males in the year 1998–1999 to the rise of up to 265 cases undergoing the implant in 2006–2007. The computed tomography (CT) imaging data are affected by the metallic implants due to the extensive range of Hounsfield units (HU) associated with erroneous CT values and cause artifacts in the CT images for radiation treatment planning [5].

The artifacts caused by the prostheses interfere with the contouring of the target and organs, and the implants also complicate the radiation planning process due to incorrectly modeled radiation dose scattering and dose attenuation, leading to dose uncertainty [6]. Similar problems have been encountered in conventional planning for external beam radiation therapy with IMRT and VMAT techniques [7]. However, there have been measures to overcome the same. The American Association of Physicists in Medicine Task Group (AAPM-TG) report 63 has provided a few solutions for managing treatment planning with the hip implant in place [6]. Also, it is not easy to assess the prosthesis electron density accurately. Thus, the planning systems find it difficult to accurately model the dose in, near, and beyond the implant for the megavoltage photon beam placed [6].

With the 10-MV treatment beam, the attenuation occurring due to the titanium hip implant is about 2.5 times larger than the bone and nearly four times higher than the water-equivalent tissue [8]. Depending on the photon beam energy and hip prosthesis composition, the dose attenuation could range from 10 to 60% [6]. The backscatter increases with the increasing power and the atomic number of the material used for the prosthesis, and it becomes a significant concern at the bone-implant surface interface [9].

With the increasing radiation treatment delivery with SBRT, conventional planning has reduced [10], and hence, these measures need to be reviewed to be employed in moderate to ultra-hypofractionation, including SBRT. The recent CHHip trial also reported hesitancy in treating patients with bilateral hip implants with SBRT [11]. The hip implants are not considered contraindications for CT and MRI simulation or radiation fraction delivery [12]. In this review article, we aim to present the technical aspects associated with planning the SBRT cases with bilateral hip implants and report the accurate and safe approach for prostate radiation treatment in the presence of bilateral hip implants.

Methodology

We performed a PubMed search with the following MesH terms: Prostate Cancer, Prostate treatment, Prostatic malignancy AND SBRT, SABR, VMAT, IMRT, AND bilateral hip implants, bilateral hip prosthesis, metallic hip implants, along with the articles published in the various oncology journals through google search and the individual websites. The literature search was focused between January 1990 to June 2022. The focus was on the best available practice regarding radiation therapy delivery in institutes that addressed the treatment planning and delivery of bilateral hip implants in the patient population with prostate cancer.

Results and discussion

Imaging for target delineation/planning

Image registration

Due to image artifacts, hip implants can cause problems in defining the clinical target volume (CTV) of the prostate and the organs at risk, including the bladder and the rectum. The role of MRI in such scenarios is valuable, and the combined CT-MR image registration helps define the target and the organs at risk better [12]. The CT and MR images need to be transferred to the treatment planning system for image registration, which is done using at least three bony landmarks, and after that, proceed with planning. Other measures that deal with the problem of target delineation include approximating target volumes from a predicted anatomical position using the available bone landmarks. However, this does not provide the conformality needed for SBRT planning. The CT-MR registration has shown improvement in the target delineation. Studies have shown that CT-derived prostate volumes are more extensive than MR-derived volumes, especially toward the seminal vesicles (by 6 mm) and the apex of the prostate [13]. Similarly, it was noted that the dose-volume histograms from CT and MRI comparison showed we could spare a mean of 10% of rectal volume and approximately 5% of bladder and femoral heads, respectively; thus, the co-registration can help avoid the overestimation of CTV with CT images compared with MRI [14]. The concern lies in the CT-based imaging on the flat couch and the MR images undertaken on a standard MR couch with a firm mattress. However, the difference did not impact the registration and further did not impact the target and organs at-risk delineation [12]. The study also noted on qualitative analysis that the phantom-filled water tubes demonstrated distortions less than 0.2 cm at distances greater than 3 cm from the prosthesis.

Metal artifact reduction (MAR) algorithms for CT imaging

Metal artifact reduction (MAR) methods have been extensively reviewed [15], and commercial MAR helps improve treatment planning. It is also easy to use as it requires minimal skills for the operations; for example, the ker-MAR has been shown to have improved efficiency in head and neck cases [16]. On the other hand, the MRI-based CT MAR [17] requires the manual selection of CT scans for artifact correction. Similarly, the deep learning-based algorithms (DL-MAR) do not require paired data for artifact corrections with similar outcomes compared to the water density override. The commercial MAR methods, including O-MAR, iMAR, Smart MAR, and SEMAR, are commonly used and have the advantage of a wide range of clinical applications. However, they have shown incomplete artifact removal with Hounsfield unit errors and may induce new artifacts [18,19,20]. The density correction methods, though available in the treatment planning system, require expertise and training for artifact identification and density overrides [21, 22]. The deep-learning-based MAR algorithms do not require sinogram or paired data and help reduce the metal artifacts on the CT scans with comparable performance to density correction for dose calculation accuracy. However, the drawback is that it requires paired data and depends upon the NMAR application [23,24,25]. Considering the advantages and the disadvantages of the MAR method, it appears to be a promising solution for developing these algorithms, which would help target specific patterns of metal artifacts and specific anatomical structures, thus proving beneficial for future radiation treatment applications. The commercially available MAR options are shown in Table 1. Figure 1 shows an axial contrast-enhanced CT scan of a 76-year-old man with a right metal hip prosthesis who underwent radiation treatment for intermediate-risk prostate cancer (A) and the imaging on quality assurance (B). Figure 2 shows (A) Standard filtered back projection reconstruction of a phantom with metal implants removed (B) standard filtered back projection reconstruction of a phantom with metal implants in place (C) MAR reconstruction with metal implants in place. Table 2 summarizes various publication results for imaging for target delineation and planning with hip prostheses.

Table 1 Various MAR Applications available for RT Applications
Table 2 Summary of measures and results for imaging for target delineation and planning with hip prostheses

Use of multi-energy CT for artifact reduction

The CT systems are equipped with polychromatic X-rays comprising a spectrum of varying energy photons representing the cumulative transmitted intensity at each energy forming the complete beam. Using such beams for longer path lengths and high atomic number materials can cause beam hardening artifacts. When the correction algorithms are unmet, they can lead to shading and dark bands on the planning images [30]. The multi-energy CT scans can help improve beam-hardening correction using virtual monoenergetic images used in CT myocardial perfusion applications [31], and similar improvements in the images have been found in dental restorations [32]. Virtual monoenergetic images help reduce image artifacts. In one of the studies, it was recommended to use the routine reconstruction of the virtual monoenergetic images at 50 keV using the dual-energy CT, and for reducing the beam-hardening artifacts, the virtual monoenergetic imaging at 120 keV was found to be useful [33]. Other studies have shown a similar impact in reducing the metal artifacts, especially with the prosthesis in situ [26,27,28, 34,35,36].

Treatment planning with Megavoltage CT (MVCT) images of helical tomotherapy

Tomotherapy generates lower energy MVCT images using the same therapeutic megavoltage x-ray beam, and the uniformity and spatial resolution of these MVCT images are comparable to the diagnostic CT images [37]. These MVCT images have poorer noise and low-contrast resolution outcomes than diagnostic CT scanners but produce sufficient contrast to help delineate soft tissue structures. A case study of prostate cancer with bilateral hip implants reported the use of MVCT with helical tomotherapy showed no artifacts in the MVCT, and the same images were used for the planning, and the plan was optimized to the planned constraints [38]. The reported patient image dose using MVCT with helical tomotherapy is 1.5 cGy per scan [39], and the use of MV-CBCT to complement standard CT for target definition in prostate cancers being treated with hip implants has been reported [29].

Ultrasound used for target volume definition

Using stereotactic ultrasound for daily prostate image guidance has helped reduce the intrafraction motion during the treatment [40, 41] and define the PTV [42]. A case report utilizing the definition of PTV based on the ultrasound images acquired during the CT simulation and matched to the planning CT and the ultrasound referenced to the isocenter of the CT simulator showed better matching of the prostate, seminal vesicles, and the target volumes [43]. The prosthesis-associated artifacts can be further reduced when added with advanced reconstruction algorithms with artifact suppression [44, 45] and dual-energy CT [46].

Treatment planning

IMRT planning in bilateral hip prosthesis

The reports on using IMRT while planning in the presence of hip implants have employed dosimetric constraints focusing on no dose entering or exiting the prosthesis by using non-coplanar beams [47]. Another study utilized coplanar beams and avoided the prosthesis in the beam’s eye view and found IMRT to be superior in terms of target coverage and reduction in doses to organs at risk with a significant decrease in the rectum dose with composite V80% of 35% for IMRT vs. 70% for the 3D-CRT plan and bladder dose reduced to 9% vs. 20% [48].

The AAPM TG 63 recommends avoiding beams passing through the hip prostheses [6] based on results from previous studies that demonstrated an impact on the dosimetric outcomes [49]. The advantage is that the certainty for the PTV dose is achieved, and there is a possibility to expand the gantry angles for coplanar IMRT beams, leading to improved dose-volume histograms as compared to that without the modification related to the implant, causing the beams to be more constrained to narrow anterior and posterior range [6]. In one case study inverse IMRT planning with geometric constraint instead of the dosimetric constraint was employed [50], defying the usual understanding proposed by AAPM TG 63 to avoid beams passing through hip prostheses that need dosimetric constraints. They concluded that avoiding beam entry alone would provide certainty of PTV dose and expand the possible gantry angles for coplanar IMRT beams. The beam attenuation while passing through the patient and the ability of the beams to fan out over more angles would be made possible by constraining beams to avoid the prosthesis, thereby causing a low dose to the prosthesis itself and the other normal tissue without compromising the target coverage. The drawback associated with this technique is related to the isocenter shifts to target implanted fiducial markers, which leads to the defined constraints for the beam meant for the implant shielding to be violated [51].

Another study employed the use of the IMRT-PAV (Prosthesis avoidance volumes) technique wherein the beam setup default without prosthesis in place is used as the starting point, and adjustments are made possible with the beam’s eye projection of the planning target volume (PTV) [52]. The beams could deposit the dose in the prosthetic device at the exit site of the PTV as dose delivery through the implant device at the entrance site of the PTV is prevented. This was achieved by defining a virtual organ-at-risk for each beam passing through the implant. These avoidance volumes needed to be placed outside the patient at minimal distances from the body contour and avoid overlaps with the beam’s eye view projections of other beams. The avoidance volumes were then delineated in the transverse images, conforming the view projections in front of the PTV with an additional 6 mm margin to account for the set-up uncertainties. The projection of each avoidance volume was evaluated after assigning each avoidance volume the electron density of water using quadratic overdose constraint for each avoidance volume and restricting the root mean square dose excess to 0.02 Gy. The results showed dose delivery to the surrounding bone around the prosthesis was higher, but the maximum dose was within constraints, and mean amounts to the bladder and rectum were reduced by up to 25%. The technique required more time for delineation, but the time for treatment planning was decreased due to applying the default beam setup.

The comparison of the coplanar and non-coplanar IMRT planning using beams that entirely avoided the implant but did not always cover the entire PTV showed that the non-coplanar plan had better rectum sparing; however, larger bladder volume was exposed to doses up to 50 Gy [53]. The study involved using 11 IMRT beams at three couch angles, resulting in a significantly longer treatment time than seven coplanar sets of treatment beams with gantry angles close to the beam directions that would be used without the prosthesis [52]. The use of Monte Carlo calculations was not found to be essential for obtaining the dose distributions as none of the photons that first traversed the prosthesis delivered the primary dose to the target or the organs at risk, and the difference in the dose delivered was within 2% or 2 mm in regions with high-dose gradients [54].

Another technique has been described that involved excluding beamlets from the optimization that would deliver the dose to the PTV while first passing through the implant [55], which led to the reduction in dose delivery to the bladder and rectum with an enhanced degree of freedom in selecting beamlets with predominantly lateral PTV irradiation. The rectal sparing was better achieved using 11–13 coplanar beams. However, there was a note of increased patient volume being treated to low doses and increased treatment times due to setting up times and each beam verification. Longer treatment times have been shown to impact the intrafraction motion [56], and thus, the treatment time could be reduced by utilizing nine coplanar beams with 10MV photon beams [55].

VMAT planning strategies with bilateral hip prosthesis

VMAT planning has become a standard of treatment delivery in prostate cancer [57, 58]. However, prosthesis avoidance is difficult due to several partial arcs needed for the beam arrangements. However, the delivery of treatment with the VMAT technique is possible with the use of the avoidance sector feature with the Varian linear accelerators, leading to the use of a complete arc and the beam shutting off temporarily at gantry angles that could deliver entry dose to the hip prosthesis [59, 60]. Another study utilized complete arcs but limited the hip prosthesis entry dose with an avoidance optimization structure around the implant with a 1 cm margin and a dose constraint of 500 cGy [61]. There has been a comparison of three different VMAT planning techniques, including two complete arcs with a maximum dose of 500 cGy to the prosthesis, two full arcs with a dose maximum of 500 cGy to 105 of the prosthesis, and six partial arcs avoiding the prosthesis and found improved dose-volume metrics for the rectum and bladder with us of the full arcs and significant improvement in the conformality, homogeneity and gradient indices favoring the use of the full arcs and improved dosimetric quality and deliverability with minimal entrance doses to each prosthesis [62].

Acuros XB algorithm and comparison with Monte Carlo calculations

The Acuros XB (AXB) algorithm by Varian Medical Systems, Inc. (VMS) (Palo Alto, CA) is a grid-based linear Boltzmann transport equation solver as a part of the Eclipse treatment planning system (TPS) (VMS) and was the first to help sharpen the dose gradients in the vicinity of the metallic implant [63, 64]. Studies have shown that the AXB algorithm can produce dose calculations with accuracy close to the Monte Carlo model [63], and the algorithm has been compared with the Monte Carlo regarding the lateral dose profile [64]. A comparative analysis of the use of the AXB on clinical patient plans showed the algorithm was reliable as a calculation algorithm in the case of planning in the presence of hip implants. It calculated the correct dose distribution at the metal-target interface. Thus, it helps avoid the need for the beams to traverse through the implant [65]. The authors also concluded that dose distribution calculations by the AXB algorithm showed minor deviations, on average within 2.5%, when compared to measurements and the MC model, but the performance was considered acceptable. An anisotropic analytical algorithm (AAA) was discouraged due to the uncertainty associated with the implant interface and inaccurate accounting of the backscatter due to the implant and the prosthesis-associated beam attenuation.

Target localization during daily treatment

CBCT for daily image guidance

CBCT helps improve tumor targeting during delivery of the radiation treatment, helps reduce the dose to the organs at risk, and helps in the correct patient setup while monitoring with adequate soft-tissue visualization [66]. However, these series do not provide accurate Hounsfield units and do not have value for direct dose calculation [67]. However, with the hip prostheses in place, the CBCT images show much more scatter than the conventional fan-beam-based CT images [68]. Thus, Hounsfield unit calibration is required. It resulted in dose-calculation errors of > 5% compared with planning CT-based dose calculations when Hounsfield units-electron density conversion curves were based on average CBCT HU values for different treatment sites [69]. Other measures to adjust the Hounsfield unit include mapping these units in the CT images to equivalent points in the CBCT image geometry after deformable image registration (DIR) [70] or by using image cumulative histograms to adjust the units for the CBCT and the planning CT images [67]. The use of fat and muscles instead of water can also be employed and has been shown to have resulted in a dosimetric difference of < 2% [71] and has been referred to as density override. In one study, the CBCT voxels were assigned as water alone, followed by water or bone, then compared for scatter correction. The automated density override approach showed six different densities (air, lung, adipose, connective tissue, bone, and higher densities for prosthesis), and the CBCT image histograms were binned into six density levels. The comparison with the planning CT images showed superior outcomes in more minor patients with an anteroposterior diameter of less than 25 cm [72]. Previous studies have shown the need for a multilevel threshold algorithm to be helpful with single hip prosthesis and have helped reduce the operator time [73].

Electromagnetic localization tracking accuracy

The implantable electromagnetic transponders help continuously monitor the prostate location and help reduce the PTV margins, which helps reduce the doses to organs at risk, particularly the anterior rectal wall [74]. Generally, using these transponders has been considered a contraindication due to the interference of the radiofrequency signals of the transponders interfering with the prosthesis [75]. However, Bittner et al., in their case report, evaluated the reliability of positional data obtained through the tracking of the prostate phantom in the presence of a hip prosthesis and concluded that there was minimal effect on the ability to track these transponders [76] accurately. The maximum offset and the average offset recorded with the unilateral hip implant were 0.7 mm and 0.5 mm, and the bilateral implants were 0.5 mm and 0.3 mm, respectively.

Monitoring intrafraction motion

With the increased use of image-guided radiation therapy, the need for tumor motion management has been explored [77, 78] due to prostate motion associated with rectal and bladder filling differences. With the use of software related to commercial tracking systems, including Calypso (Varian Medical Systems, Palo Alto, CA) and Micropos (Micropos Medical AB, Goteborg, Sweden) [79,80,81,82,83], the implanted fiducials can be visualized and help assess the intrafraction motion. Also, the automated beam hold (ABH) provided with the TrueBeam linear accelerator (version 2.0 and onward, Varian Medical Systems) helps assess the kilovoltage images acquired during the beam on time. These images are the triggered images, which are obtained at predefined intervals during treatment delivery, and the implanted fiducials are automatically visualized [84]. The use of ABH accurately detects increased tumor motion, thereby increasing the prostate treatment delivery accuracy. Rosario et al. showed successful intrafraction motion detection with kV imaging, which helped reduce the PTV margins from 6 mm to 5 mm [84]. Similar results were seen with 3-dimension imaging using the Calypso system and function of time, and they showed that for treatments completed in 5 min (VMAT) and 10 min (IMRT), the proportion for the prostate to shift by > 3 mm was 4% and 12%, respectively [85]. The role is thus under evaluation and needs to be investigated further to incorporate the need for treatment planning for prostate SBRT in the presence of the hip prosthesis.

Conclusion

In our review, we have presented the image-registration techniques, contouring, and planning aspects in patients with prostate cancer in the presence of hip implants. There is no available retrospective or prospective literature providing the clinical outcomes of SBRT in prostate cancer in the presence of hip implants, highlighting the fact that there is a hindrance in considering such patients for treatment with SBRT. The mentioned literature in this review looks into dosimetric aspects, and some studies have been conducted on the cadavers. Hence, the literature available is not robust enough to conclude whether the measures highlighted in the review can be implemented in the real-case scenario, and further prospective studies are needed to conclusively state whether such hip prosthesis is necessarily a contraindication for SBRT. With a rise in the number of patients undergoing hip replacement surgeries and the incidence of prostate cancer also rising, there is a need for better clarity on the aspects of SBRT treatment. There is a concern regarding the maximum dose in the bladder and the rectum, target delineation, image matching during the treatment, and transponders matching. However, the available reports encourage planning patients with this treatment modality. The review also highlights the need for future clinical trials to look into the survival and toxicity profiles on the impact of the prosthesis’s presence in prostate cancer management. Overall, there does not seem to be a need to consider the presence of hip implants as a contraindication for the SBRT treatment.

Fig. 1
figure 1

Axial contrast-enhanced CT scans of a 76-year-old man with a right metal hip prosthesis who underwent radiation treatment for intermediate-risk prostate cancer (A) and the imaging on quality assurance (B)

Fig. 2
figure 2

(A) standard filtered back projection reconstruction of a phantom with metal implants removed. (B) standard filtered back projection reconstruction of a phantom with metal implants in place. (C) MAR reconstruction with metal implants in place

Data availability

Not applicable.

References

  1. Cihan Y. The role and importance of SBRT in prostate cancer. Int Braz J Urol. 2018 Nov-Dec;44(6):1272–4. https://doi.org/10.1590/S1677-5538.IBJU.2018.0484. PMID: 30325604; PMCID: PMC6442169.

  2. Aghdam N, Pepin A, Buchberger D, Hirshberg J, Lei S, Ayoob M, Danner M, Yung T, Kumar D, Collins BT, Lynch J, Kataria S, Suy S, Collins SP. Stereotactic body Radiation Therapy (SBRT) for prostate Cancer in men with a high baseline international prostate symptom score (IPSS ≥ 15). Front Oncol. 2020;10:1060. https://doi.org/10.3389/fonc.2020.01060. PMID: 32719744; PMCID: PMC7350884.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Maradit Kremers H, Larson DR, Crowson CS, Kremers WK, Washington RE, Steiner CA, Jiranek WA, Berry DJ. Prevalence of total hip and knee replacement in the United States. J Bone Joint Surg Am. 2015;97(17):1386–97. https://doi.org/10.2106/JBJS.N.01141. PMID: 26333733; PMCID: PMC4551172.

    Article  PubMed  Google Scholar 

  4. Population Health Division. The Health of the people of New South Wales – Report of the Chief Health Officer, Data Book – Burden of Disease. Sydney: NSW Department of Health; 2008.

    Google Scholar 

  5. Banaee N, Barough MS, Asgari S, Hosseinzadeh E, Salimi E. Evaluating the effects of metal artifacts on dose distribution of the pelvic region. J Cancer Res Ther 2021 Apr-Jun;17(2):450–4. https://doi.org/10.4103/jcrt.JCRT_786_19.

  6. Reft C, Alecu R, Das IJ, Gerbi BJ, Keall P, Lief E, Mijnheer BJ, Papanikolaou N, Sibata C, Van Dyk J, AAPM Radiation Therapy Committee Task Group 63. Dosimetric considerations for patients with HIP prostheses undergoing pelvic irradiation. Report of the AAPM Radiation Therapy Committee Task Group 63. Med Phys. 2003;30(6):1162–82. https://doi.org/10.1118/1.1565113.

    Article  PubMed  Google Scholar 

  7. Giantsoudi D, De Man B, Verburg J, Trofimov A, Jin Y, Wang G, Gjesteby L, Paganetti H. Metal artifacts in computed tomography for radiation therapy planning: dosimetric effects and impact of metal artifact reduction. Phys Med Biol. 2017;62(8):R49–80. https://doi.org/10.1088/1361-6560/aa5293. Epub 2017 Mar 21.

    Article  PubMed  Google Scholar 

  8. Attix FH. Introduction to radiological phyfsics and radiation dosimetry. New York: Wiley; 1986.

    Book  Google Scholar 

  9. Das IJ, Kahn FM. Backscatter dose perturbation at high atomic number interfaces in megavoltage photon beams. Med Phys. 1989;16(3):367–75. https://doi.org/10.1118/1.596345.

    Article  CAS  PubMed  Google Scholar 

  10. Mahase SS, D’Angelo D, Kang J, Hu JC, Barbieri CE, Nagar H. Trends in the use of stereotactic body radiotherapy for treatment of prostate Cancer in the United States. JAMA Netw Open. 2020;3(2):e1920471. https://doi.org/10.1001/jamanetworkopen.2019.20471. Erratum in: JAMA Netw Open. 2020;3(10):e2027727.

    Article  PubMed  Google Scholar 

  11. Dearnaley D, Syndikus I, Mossop H, Khoo V, Birtle A, Bloomfield D, Graham J, Kirkbride P, Logue J, Malik Z, Money-Kyrle J, O’Sullivan JM, Panades M, Parker C, Patterson H, Scrase C, Staffurth J, Stockdale A, Tremlett J, Bidmead M, Mayles H, Naismith O, South C, Gao A, Cruickshank C, Hassan S, Pugh J, Griffin C, Hall E. CHHiP Investigators. Conventional versus hypofractionated high-dose intensity-modulated radiotherapy for prostate cancer: 5-year outcomes of the randomised, non-inferiority, phase 3 CHHiP trial. Lancet Oncol. 2016;17(8):1047–1060. doi: 10.1016/S1470-2045(16)30102-4. Epub 2016 Jun 20. Erratum in: Lancet Oncol. 2016;17 (8):e321.

  12. Charnley N, Morgan A, Thomas E, Wilson S, Bacon S, Wilson D, Bottomley D. The use of CT-MR image registration to define target volumes in pelvic radiotherapy in the presence of bilateral hip replacements. Br J Radiol. 2005;78(931):634–6. https://doi.org/10.1259/bjr/28412864.

    Article  CAS  PubMed  Google Scholar 

  13. Rasch C, Barillot I, Remeijer P, Touw A, van Herk M, Lebesque JV. Definition of the prostate in CT and MRI: a multi-observer study. Int J Radiat Oncol Biol Phys. 1999;43(1):57–66. https://doi.org/10.1016/s0360-3016(98)00351-4.

    Article  CAS  PubMed  Google Scholar 

  14. Sannazzari GL, Ragona R, Ruo Redda MG, Giglioli FR, Isolato G, Guarneri A. CT-MRI image fusion for delineation of volumes in three-dimensional conformal radiation therapy in the treatment of localized prostate cancer. Br J Radiol. 2002;75(895):603–7. https://doi.org/10.1259/bjr.75.895.750603.

    Article  CAS  PubMed  Google Scholar 

  15. Puvanasunthararajah S, Fontanarosa D, Wille ML, Camps SM. The application of metal artifact reduction methods on computed tomography scans for radiotherapy applications: a literature review. J Appl Clin Med Phys. 2021;22(6):198–223. https://doi.org/10.1002/acm2.13255.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Nielsen JS, Van Leemput K, Edmund JM. MR-based CT metal artifact reduction for head‐and‐neck photon, electron, and proton radiotherapy. Med Phys. 2019;46:4314–23.

    Article  PubMed  Google Scholar 

  17. Park PC, Schreibmann E, Roper J, Elder E, Crocker I, Fox T, Zhu XR, Dong L, Dhabaan A. MRI-based computed tomography metal artifact correction method for improving proton range calculation accuracy. Int J Radiat Oncol Biol Phys. 2015;91(4):849 – 56. https://doi.org/10.1016/j.ijrobp.2014.12.027. PMID: 25752400.

  18. Akdeniz Y, Yegingil I, Yegingil Z. Effects of metal implants and a metal artifact reduction tool on calculation accuracy of AAA and Acuros XB algorithms in small fields. Med Phys. 2019;46:5326–35.

    Article  PubMed  Google Scholar 

  19. Jia Y, Zhao L, Cheng C, Mcdonald M, Das I. SU-E‐T‐450: dose perturbation of spinal metal implants in proton beam therapy. Med Phys. 2013;40:309.

    Article  Google Scholar 

  20. Righetto R, Clemens LP, Lorentini S, Fracchiolla F, Algranati C, Tommasino F, Dionisi F, Cianchetti M, Schwarz M, Farace P. Accurate proton treatment planning for pencil beam crossing titanium fixation implants. Phys Med. 2020;70:28–38. Epub 2020 Jan 15. PMID: 31954210.

    Article  PubMed  Google Scholar 

  21. Acquah GF, Kyeremeh PO, Hasford F, Boadu M, Sosu EK, Inkoom S. Evaluation of metallic implant artifact on photon beam calculation algorithms using a CIRS thorax phantom. J Radiat Res Appl Sci. 2018;11:347–9.

    CAS  Google Scholar 

  22. Parenica HM, Mavroidis P, Jones W, Swanson G, Papanikolaou N, Stathakis S. VMAT optimization and dose calculation in the presence of metallic hip prostheses. Technol Cancer Res Treat. 2019;18:1–10.

    Article  Google Scholar 

  23. Huang X, Wang J, Tang F, Zhong T, Zhang Y. Metal artifact reduction on cervical CT images by deep residual learning 08 information and Computing sciences 0801 Artificial Intelligence and Image Processing. Biomed Eng Online. 2018;17:1–15.

    Google Scholar 

  24. Gjesteby L, Shan H, Yang Q, Xi Y, Jin Y, Giantsoudi D, Paganetti H, De Man B, Wang G. A dual-stream deep convolutional network for reducing metal streak artifacts in CT images. Phys Med Biol. 2019;64(23):235003. https://doi.org/10.1088/1361-6560/ab4e3e. PMID: 31618724.

  25. Koike Y, Anetai Y, Takegawa H, Ohira S, Nakamura S, Tanigawa N. Deep learning-based metal artifact reduction using cycle‐consistent adversarial network for intensity‐modulated head and neck radiation therapy treatment planning. Phys Med. 2020;78:8–14.

    Article  PubMed  Google Scholar 

  26. Filograna L, Magarelli N, Leone A, Guggenberger R, Winklhofer S, Thali MJ, Bonomo L. Value of monoenergetic dual-energy CT (DECT) for artefact reduction from metallic orthopedic implants in post-mortem studies. Skeletal Radiol. 2015;44(9):1287–94. https://doi.org/10.1007/s00256-015-2155-z. Epub 2015 May 12. PMID: 25962510.

    Article  PubMed  Google Scholar 

  27. Guggenberger R, Winklhofer S, Osterhoff G, Wanner GA, Fortunati M, Andreisek G, Alkadhi H, Stolzmann P. Metallic artefact reduction with monoenergetic dual-energy CT: systematic ex vivo evaluation of posterior spinal fusion implants from various vendors and different spine levels. Eur Radiol. 2012;22(11):2357–64. https://doi.org/10.1007/s00330-012-2501-7. Epub 2012 May 30. PMID: 22645043.

    Article  CAS  PubMed  Google Scholar 

  28. Yue D, Fan Rong C, Ning C, Liang H, Ai Lian L, Ru Xin W, Ya Hong L. Reduction of metal artifacts from unilateral hip arthroplasty on dual-energy CT with metal artifact reduction software. Acta Radiol. 2018;59(7):853–60. Epub 2017 Sep 12. PMID: 28899125.

    Article  PubMed  Google Scholar 

  29. Aubin M, Morin O, Chen J, Gillis A, Pickett B, Aubry JF, Akazawa C, Speight J, Roach M 3rd, Pouliot J. The use of megavoltage cone-beam CT to complement CT for target definition in pelvic radiotherapy in the presence of hip replacement. Br J Radiol. 2006;79(947):918–21. https://doi.org/10.1259/bjr/19559792. Epub 2006 Aug 17.

    Article  CAS  PubMed  Google Scholar 

  30. McCollough CH, Boedeker K, Cody D, Duan X, Flohr T, Halliburton SS, Hsieh J, Layman RR, Pelc NJ. Principles and applications of multienergy CT: Report of AAPM Task Group 291. Med Phys. 2020;47(7):e881-e912. doi: 10.1002/mp.14157. Epub 2020 May 28. Erratum in: Med Phys. 2021;48(5):2694. PMID: 32215937.

  31. Rodriguez-Granillo GA, Carrascosa P, Cipriano S, De Zan M, Deviggiano A, Capunay C, Cury RC. Beam hardening artifact reduction using dual energy computed tomography: implications for myocardial perfusion studies. Cardiovasc Diagn Ther. 2015;5(1):79–85. https://doi.org/10.3978/j.issn.2223-3652.2015.01.13.

    Article  PubMed  PubMed Central  Google Scholar 

  32. Stolzmann P, Winklhofer S, Schwendener N, Alkadhi H, Thali MJ, Ruder TD. Monoenergetic computed tomography reconstructions reduce beam hardening artifacts from dental restorations. Forensic Sci Med Pathol. 2013;9(3):327–32. https://doi.org/10.1007/s12024-013-9420-z. Epub 2013 Mar 20.

    Article  PubMed  Google Scholar 

  33. Albrecht MH, Vogl TJ, Martin SS, Nance JW, Duguay TM, Wichmann JL, De Cecco CN, Varga-Szemes A, van Assen M, Tesche C, Schoepf UJ. Review of clinical applications for virtual Monoenergetic Dual-Energy CT. Radiology. 2019;293(2):260–71. https://doi.org/10.1148/radiol.2019182297. Epub 2019 Sep 10.

    Article  PubMed  Google Scholar 

  34. Bamberg F, Dierks A, Nikolaou K, Reiser MF, Becker CR, Johnson TR. Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation. Eur Radiol. 2011;21:1424–9.

    Article  PubMed  Google Scholar 

  35. Cha J, Kim HJ, Kim ST, Kim YK, Kim HY, Park GM. Dual-energy CT with virtual monochromatic images and metal artifact reduction software for reducing metallic dental artifacts. Acta Radiol. 2017;58:1312–9.

    Article  PubMed  Google Scholar 

  36. Dunet V, Bernasconi M, Hajdu SD, Meuli RA, Daniel RT, Zerlauth JB. Impact of metal artifact reduction software on image quality of gemstone spectral imaging dual-energy cerebral CT angiography after intracranial aneurysm clipping. Neuroradiology. 2017;59:845–52.

    Article  PubMed  Google Scholar 

  37. Meeks SL, Harmon JF Jr, Langen KM, Willoughby TR, Wagner TH, Kupelian PA. Performance characterization of megavoltage computed tomography imaging on a helical tomotherapy unit. Med Phys. 2005;32(8):2673–81. https://doi.org/10.1118/1.1990289.

    Article  PubMed  Google Scholar 

  38. Alongi F, Fodor A, Maggio A, Cozzarini C, Fiorino C, Broggi S, Alongi P, Calandrino R, Muzio ND. Megavoltage CT images of helical tomotherapy unit for radiation treatment simulation: impact on feasibility of treatment planning in a prostate cancer patient with bilateral femoral prostheses. Tumori. 2011 Mar-Apr;97(2):221–4. https://doi.org/10.1177/030089161109700215.

  39. Shah AP, Langen KM, Ruchala KJ, Cox A, Kupelian PA, Meeks SL. Patient dose from megavoltage computed tomography imaging. Int J Radiat Oncol Biol Phys. 2008;70(5):1579–87. Epub 2008 Jan 30.

    Article  PubMed  Google Scholar 

  40. Fuss M, Cavanuagh SX, Fuss C, Cheek DA, Salter BJ. Daily stereotactic ultrasound prostate targeting: inter-user variability. Technol Cancer Res Treat. 2003;2:161–70.

    Article  PubMed  Google Scholar 

  41. Boda-Heggemann J, Köhler FM, Küpper B, Wolff D, Wertz H, Mai S, Hesser J, Lohr F, Wenz F. Accuracy of ultrasound-based (BAT) prostate-repositioning: a three-dimensional on-line fiducial-based assessment with cone-beam computed tomography. Int J Radiat Oncol Biol Phys. 2008;70(4):1247-55. https://doi.org/10.1016/j.ijrobp.2007.12.003. PMID: 18313531.

  42. Dube F, Mahadevan A, Sheldon T. Fusion of CT and 3D ultrasound for prostate delineation of patients with metallic hip prostheses. Int J Radiat Oncol Biol Phys. 2009;75:327.

    Article  Google Scholar 

  43. Boda-Heggemann J, Haneder S, Ehmann M, Sihono DSK, Wertz H, Mai S, Kegel S, Heitmann S, von Swietochowski S, Lohr F, Wenz F. Stereotactic ultrasound for target volume definition in a patient with prostate cancer and bilateral total hip replacement. Pract Radiat Oncol. 2015 May-Jun;5(3):197–202. Epub 2014 Sep 26.

  44. Prell D, Kyriakou Y, Kachelrie M, Kalender WA. Reducing metal artifacts in computed tomography caused by hip endoprostheses using a physics-based approach. Invest Radiol. 2010;45:747–54.

    Article  PubMed  Google Scholar 

  45. Robertson DD, Yuan J, Wang G, Vannier MW. Total hip prosthesis metal-artifact suppression using iterative deblurring reconstruction. J Comput Assist Tomogr. 1997;21:293–8.

    Article  CAS  PubMed  Google Scholar 

  46. Lewis M, Reid K, Toms AP. Reducing the effects of metal artefact using high keV monoenergetic reconstruction of dual energy CT (DECT) in hip replacements. Skeletal Radiol. 2013;42:275–82.

    Article  PubMed  Google Scholar 

  47. Brooks C, Cheung RM, Kudchadker RJ. Intensity-modulated radiation therapy with noncoplanar beams for treatment of prostate cancer in patients with bilateral hip prosthesis-a case study. Med Dosim. 2010 Summer;35(2):87–91. Epub 2009 Apr 18. PMID: 19931019.

  48. Su A, Reft C, Rash C, Price J, Jani AB. A case study of radiotherapy planning for a bilateral metal hip prosthesis prostate cancer patient. Med Dosim. 2005 Fall;30(3):169 – 75. https://doi.org/10.1016/j.meddos.2005.06.001.

  49. Hazuka MB, Ibbott GS, Kinzie JJ. Hip prostheses during pelvic irradiation: effects and corrections. Int J Radiat Oncol Biol Phys. 1988;14(6):1311–7. https://doi.org/10.1016/0360-3016(88)90412-9.

    Article  CAS  PubMed  Google Scholar 

  50. Martin DA, Hruby G, Whitaker MK, Foo KY. Constrained-beam inverse planning for intensity-modulated radiation therapy of prostate cancer patients with bilateral hip prostheses. J Med Imaging Radiat Oncol. 2012;56(6):703–7. https://doi.org/10.1111/j.1754-9485.2012.02456.x. Epub 2012 Oct 9.

    Article  PubMed  Google Scholar 

  51. Hwang AB, Kinsey E, Xia P. Investigation of the dosimetric accuracy of the isocenter shifting method in prostate cancer patients with and without hip prostheses. Med Phys. 2009;36(11):5221–7.

    Article  PubMed  Google Scholar 

  52. van der Est H, Prins P, Heijmen BJ, Dirkx ML. Intensity modulated radiation therapy planning for patients with a metal hip prosthesis based on class solutions. Pract Radiat Oncol. 2012 Jan-Mar;2(1):35–40. https://doi.org/10.1016/j.prro.2011.04.006. Epub 2011 Jun 8.

  53. Brooks C, Cheung RM, Kudchadker RJ. Intensity-modulated radiation therapy with noncoplanar beams for treatment of prostate cancer in patients with bilateral hip prosthesis–a case study. Med Dosim. 2010;35:87–91.

    Article  PubMed  Google Scholar 

  54. Jeleń U, Alber M. A finite size pencil beam algorithm for IMRT dose optimization: density corrections. Phys Med Biol. 2007;52:617–33.

    Article  PubMed  Google Scholar 

  55. Voet PW, Dirkx ML, Breedveld S, Heijmen BJ. Automated generation of IMRT treatment plans for prostate cancer patients with metal hip prostheses: comparison of different planning strategies. Med Phys. 2013;40(7):071704. https://doi.org/10.1118/1.4808117.

    Article  PubMed  Google Scholar 

  56. Viswanathan A, Yorke E, Marks L, Eifel P, Shipley W. Radiation dose-volume effects of the urinary bladder. Int J Radiat Oncol Biol Phys. 2010;76(3):S116–22.

    Article  PubMed  PubMed Central  Google Scholar 

  57. Quan EM, Li X, Li Y, Wang X, Kudchadker RJ, Johnson JL, Kuban DA, Lee AK, Zhang X. A comprehensive comparison of IMRT and VMAT plan quality for prostate cancer treatment. Int J Radiat Oncol Biol Phys. 2012;83(4):1169–78. https://doi.org/10.1016/j.ijrobp.2011.09.015. PMID: 22704703; PMCID: PMC3805837.

    Article  PubMed  PubMed Central  Google Scholar 

  58. Ren W, Sun C, Lu N, Xu Y, Han F, Liu YP, Dai J. Dosimetric comparison of intensity-modulated radiotherapy and volumetric-modulated arc radiotherapy in patients with prostate cancer: a meta-analysis. J Appl Clin Med Phys. 2016;17(6):254–62. https://doi.org/10.1120/jacmp.v17i6.6464. PMID: 27929498; PMCID: PMC5690508.

    Article  PubMed  PubMed Central  Google Scholar 

  59. Ng WL, Brunt J, Temple S, Saipillai M, Haridass A, Wong H, Malik Z, Eswar C. Volumetric modulated arc therapy in prostate cancer patients with metallic hip prostheses in a UK centre. Rep Pract Oncol Radiother. 2015 Jul-Aug;20(4):273–7. https://doi.org/10.1016/j.rpor.2015.03.006. Epub 2015 Apr 14. PMID: 26109914; PMCID: PMC4477120.

  60. Rana S, Pokharel S. A dosimetric study of volumetric modulated arc therapy planning techniques for treatment of low-risk prostate cancer in patients with bilateral hip prostheses. South Asian J Cancer. 2014;3(1):18–21. https://doi.org/10.4103/2278-330X.126506.

    Article  PubMed  PubMed Central  Google Scholar 

  61. Prabhakar R, Kumar M, Cheruliyil S, Jayakumar S, Balasubramanian S, Cramb J. Volumetric modulated arc therapy for prostate cancer patients with hip prosthesis. Rep Pract Oncol Radiother J Gt Cancer Cent Pozn Pol Soc Radiat Oncol. 2013;18(4):209–13. https://doi.org/10.1016/j.rpor.2013.03.006.

    Article  Google Scholar 

  62. To D, Xhaferllari I, Liu M, Liang J, Knill C, Nandalur S, Gustafson G, Lack D. Evaluation of VMAT Planning strategies for prostate patients with bilateral hip prosthesis. Technol Cancer Res Treat. 2021 Jan-Dec;20:15330338211038490. https://doi.org/10.1177/15330338211038490.

  63. Failla GA, Wareing T, Archambault Y, Thompson S. Acuros XB advanced dose calculation for the eclipse treatment planning system. Palo Alto, CA: Varian Medical Systems; 2010.

    Google Scholar 

  64. Lloyd SA, Ansbacher W. Evaluation of an analytic linear boltzmann transport equation solver for high-density inhomogeneities. Med Phys. 2013;40(1):011707.

    Article  CAS  PubMed  Google Scholar 

  65. Ojala J, Kapanen M, Sipilä P, Hyödynmaa S, Pitkänen M. The accuracy of Acuros XB algorithm for radiation beams traversing a metallic hip implant - comparison with measurements and Monte Carlo calculations. J Appl Clin Med Phys. 2014;15(5):4912.

    Article  PubMed  Google Scholar 

  66. Jaffray DA, Siewerdsen JH, Wong JW, Martinez AA. Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys. 2002;53:1337–49.

    Article  PubMed  Google Scholar 

  67. Fotina I, Hopfgartner J, Stock M, SteiningerT, Ltgendorf-Caucig C, Georg D. Feasibility of CBCT-based dose calculation: comparative analysis of HU adjustment techniques. Radiother Oncol. 2012;104:249–56.

    Article  PubMed  Google Scholar 

  68. Stock M, Pasler M, Birkfellner W, HomolkaP, Poetter R, Georg D. Image quality and stability of image-guided radiotherapy(IGRT) devices: a comparative study. Radio-ther Oncol. 2009;93:1–7.

    Article  Google Scholar 

  69. Richter A, Hu Q, Steglich D, Baier K, Wilbert J, Guckenberger M, Flentje M. Investigation of the usability of conebeam CT data sets for dose calculation. Radiat Oncol. 2008;3:42. https://doi.org/10.1186/1748-717X-3-42. PMID: 19087250; PMCID: PMC2648965.

    Article  PubMed  PubMed Central  Google Scholar 

  70. Van Zijtveld M, Dirkx M, Heijmen B. Correction of cone beam CT values using a planning CT for derivation of the dose of the day. Radiother Oncol. 2007;85:195–200.

    Article  PubMed  Google Scholar 

  71. Onozato Y, Kadoya N, Fujita Y, Arai K, Dobashi S, Takeda K, Kishi K, Umezawa R, Matsushita H, Jingu K. Evaluation of on-board kV cone beam computed tomography-based dose calculation with deformable image registration using Hounsfield unit modifications. Int J Radiat Oncol Biol Phys. 2014;89(2):416–23. Epub 2014 Mar 28. PMID: 24685445.

    Article  PubMed  Google Scholar 

  72. Dunlop A, McQuaid D, Nill S, Murray J, Poludniowski G, Hansen VN, Bhide S, Nutting C, Harrington K, Newbold K, Oelfke U. Comparison of CT number calibration techniques for CBCT-based dose calculation. Strahlenther Onkol. 2015;191(12):970–8. https://doi.org/10.1007/s00066-015-0890-7. Epub 2015 Sep 24. PMID: 26403913; PMCID: PMC4656712.

    Article  PubMed  PubMed Central  Google Scholar 

  73. Almatani T, Hugtenburg R, Lewis R, Barley S, Edwards M. Simplified material assignment for cone beam computed tomography-based dose calculations of prostate radiotherapy with hip prostheses. J Radiother Pract. 2016;15:170–80.

    Article  Google Scholar 

  74. Tanyi JA, He T, Summers PA, Mburu RG, Kato CM, Rhodes SM, Hung AY, Fuss M. Assessment of planning target volume margins for intensity-modulated radiotherapy of the prostate gland: role of daily inter- and intrafraction motion. Int J Radiat Oncol Biol Phys. 2010;78(5):1579–85. Epub 2010 May 14. PMID: 20472357.

    Article  PubMed  Google Scholar 

  75. Rajendran RR, Plastaras JP, Mick R, McMichael Kohler D, Kassaee A, Vapiwala N. Daily isocenter correction with electromagnetic-based localization improves target coverage and rectal sparing during prostate radiotherapy. Int J Radiat Oncol Biol Phys. 2010;76:1092–9.

    Article  PubMed  Google Scholar 

  76. Bittner N, Butler WM, Kurko BS, Merrick GS. Effect of metal hip prosthesis on the accuracy of electromagnetic localization tracking. Pract Radiat Oncol. 2015 Jan-Feb;5(1):43 – 8. https://doi.org/10.1016/j.prro.2014.03.010. Epub 2014 May 14.

  77. Murray LJ, Lilley J, Thompson CM, Cosgrove V, Mason J, Sykes J, Franks K, Sebag-Montefiore D, Henry AM. Prostate stereotactic ablative radiation therapy using volumetric modulated arc therapy to dominant intraprostatic lesions. Int J Radiat Oncol Biol Phys. 2014;89(2):406–15. Epub 2014 Mar 28. PMID: 24685447; PMCID: PMC4018668.

    Article  PubMed  PubMed Central  Google Scholar 

  78. Lips IM, van der Heide UA, Haustermans K, van Lin EN, Pos F, Franken SP, Kotte AN, van Gils CH, van Vulpen M. Single blind randomized phase III trial to investigate the benefit of a focal lesion ablative microboost in prostate cancer (FLAME-trial): study protocol for a randomized controlled trial. Trials. 2011;12:255. https://doi.org/10.1186/1745-6215-12-255. PMID: 22141598; PMCID: PMC3286435.

    Article  PubMed  PubMed Central  Google Scholar 

  79. Adamson J, Wu Q. Prostate intrafraction motion evaluation using kV fluoroscopy during treatment delivery: a feasibility and accuracy study. Med Phys. 2008;35:1793–806.

    Article  PubMed  PubMed Central  Google Scholar 

  80. Kupelian P, Willoughby T, Mahadevan A, Djemil T, Weinstein G, Jani S, Enke C, Solberg T, Flores N, Liu D, Beyer D, Levine L. Multi-institutional clinical experience with the Calypso system in localization and continuous, real-time monitoring of the prostate gland during external radiotherapy. Int J Radiat Oncol Biol Phys. 2007;67(4):1088–98. https://doi.org/10.1016/j.ijrobp.2006.10.026. Epub 2006 Dec 21. PMID: 17187940.

    Article  PubMed  Google Scholar 

  81. Ng JA, Booth JT, Poulsen PR, Fledelius W, Worm ES, Eade T, Hegi F, Kneebone A, Kuncic Z, Keall PJ. Kilovoltage intrafraction monitoring for prostate intensity modulated arc therapy: first clinical results. Int J Radiat Oncol Biol Phys. 2012;84(5):e655–61. https://doi.org/10.1016/j.ijrobp.2012.07.2367. Epub 2012 Sep 11. PMID: 22975613; PMCID: PMC5357433.

    Article  PubMed  PubMed Central  Google Scholar 

  82. Shchory T, Schifter D, Lichtman R, Neustadter D, Corn BW. Tracking accuracy of a real-time fiducial tracking system for patient positioning and monitoring in radiation therapy. Int J Radiat Oncol Biol Phys. 2010;78:1227–34.

    Article  PubMed  Google Scholar 

  83. Kindblom J, Ekelund-Olvenmark AM, Syren H, Iustin R, Braide K, Frank-Lissbrant I, Lennernäs B. High precision transponder localization using a novel electromagnetic positioning system in patients with localized prostate cancer. Radiother Oncol. 2009;90(3):307–11. Epub 2008 Oct 24. PMID: 18952311.

    Article  PubMed  Google Scholar 

  84. Rosario T, van der Weide L, Admiraal M, Piet M, Slotman B, Cuijpers J. Toward planning target volume margin reduction for the prostate using intrafraction motion correction with online kV imaging and automatic detection of implanted gold seeds. Pract Radiat Oncol. 2018 Nov-Dec;8(6):422–8. Epub 2018 Apr 25.

  85. Tong X, Chen X, Li J, Xu Q, Lin MH, Chen L, Price RA, Ma CM. Intrafractional prostate motion during external beam radiotherapy monitored by a real-time target localization system. J Appl Clin Med Phys. 2015;16(2):5013. https://doi.org/10.1120/jacmp.v16i2.5013.

    Article  PubMed  Google Scholar 

Download references

Funding

There was no funding received for the manuscript and project to any of the authors involved.

Author information

Authors and Affiliations

Authors

Contributions

Sheen Dube: Conceptualization, methodology, writing – original draft preparationb. Vibhay Pareek: Conceptualization, methodology, writing – original draft preparationc. Mansi Barthwal: Conceptualization, methodology, writing – original draft preparationd. Febin Antony: Conceptualization, methodology, writing – original draft preparatione. David Sasaki: Supervision, Visualization, Data Curation, Writing- Reviewing and Editingf. Ryan Rivest: Supervision, Visualization, Data Curation, Writing- Reviewing and Editing.

Corresponding author

Correspondence to Vibhay Pareek.

Ethics declarations

Consent for publication

Not applicable.

Ethics approval and consent to participate

Not applicable.

Competing interests

None of the authors have any competing interests to declare.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions