Figure 5From: Finite element modeling and in vivo analysis of electrode configurations for selective stimulation of pudendal afferent fibers In vivo thresholds to evoke reflex EAS activation by intraurethral stimulation. (A) Stimulation thresholds for evoking an EAS reflex response in 11 cats with 1 mm or 2 mm ring electrodes. Inset: Reflex EAS EMG responses evoked by 1 Hz, 6 mA IES at 1 cm from the urethral meatus. (B) Normalized intensity thresholds were dependent on stimulation location (p < 0.001, one-way ANOVA), and stimulation at 2 and 3 cm required significantly higher thresholds to evoke an EAS reflex response than stimulation at all other locations (p < 0.05, Bonferroni post-hoc comparisons).Back to article page